Detection Technology camera
Default values on initialization
Op code | Operation name | Value | Explanation and default |
---|---|---|---|
| Operation Mode |
| Continuous mode |
| Gain Correction |
| Disable gain correction |
| Offset Correcting |
| Disable offset correction |
| Baseline Correcting |
| Disable baseline correction |
| PDC correction |
| Disable PDC correction |
| Pixelbinning |
| 0x00 or GUI input |
| Line averaging, the filter size is 2n |
| n = 0 |
| Output scale |
| Original |
| Offset averaging filter |
| 1 line |
| External Line Trigger Status |
| Disable external line trigger |
| External Frame Trigger Status |
| Disable external frame trigger |
| Trigger stamp parity check mode |
| Disable parity check |
| Period of heartbeat packets |
| Every 60 seconds |
| X-GCU type |
| Ethernet |
| Test pattern status |
| Disable |
| DM’s test mode |
| Disable |
| LED lights |
| Disable |
| Pixel order mode |
| Low from first, High from first |
| Energy mode |
| DE stacked |
| Offset delay mode |
| Disable offset delay time setting |
| Camera link data format |
| Normal |
After calculating calibration and geometric correction values.
| Starts/stops scanning |
| Start scanning |
Failed to obtain energy levels Error
If the energy levels are not satisfied with the timeout time according to the information in the heartbeat received every second:
Streaming Error
If no UDP packet is received for 5 seconds -> CameraErrorCode ("Camera not streaming")
Action -> Call Initialize
Heartbeat Error
If a UDP heartbeat packet is received with ErrorId ≠ 0x00
-> CameraErrorCode (Heartbeat error message)
Action -> Call Initialize
Breeze Runtime mandatory fields
Name | Min | Max | Step | Unit |
---|---|---|---|---|
| 0 | 160 | 1 | kV |
| 0 | 10 | 0.1 | mA |