Default values on initialization
|
Op code |
Operation name |
Value |
Explanation and default |
|---|---|---|---|
|
|
Operation Mode |
|
Continuous mode |
|
|
Gain Correction |
|
Disable gain correction |
|
|
Offset Correcting |
|
Disable offset correction |
|
|
Baseline Correcting |
|
Disable baseline correction |
|
|
PDC correction |
|
Disable PDC correction |
|
|
Pixelbinning |
|
0x00 or GUI input |
|
|
Line averaging, the filter size is 2n |
|
n = 0 |
|
|
Output scale |
|
Original |
|
|
Offset averaging filter |
|
1 line |
|
|
External Line Trigger Status |
|
Disable external line trigger |
|
|
External Frame Trigger Status |
|
Disable external frame trigger |
|
|
Trigger stamp parity check mode |
|
Disable parity check |
|
|
Period of heartbeat packets |
|
Every 60 seconds |
|
|
X-GCU type |
|
Ethernet |
|
|
Test pattern status |
|
Disable |
|
|
DM’s test mode |
|
Disable |
|
|
LED lights |
|
Disable |
|
|
Pixel order mode |
|
Low from first, High from first |
|
|
Energy mode |
|
DE stacked |
|
|
Offset delay mode |
|
Disable offset delay time setting |
|
|
Camera link data format |
|
Normal |
After calculating calibration and geometric correction values.
|
|
Starts/stops scanning |
|
Start scanning |
Failed to obtain energy levels Error
If the energy levels are not satisfied with the timeout time according to the information in the heartbeat received every second:
Streaming Error
If no UDP packet is received for 5 seconds -> CameraErrorCode ("Camera not streaming")
Action -> Call Initialize
Heartbeat Error
If a UDP heartbeat packet is received with ErrorId ≠ 0x00 -> CameraErrorCode (Heartbeat error message)
Action -> Call Initialize
Breeze Runtime mandatory fields
|
Name |
Min |
Max |
Step |
Unit |
|---|---|---|---|---|
|
|
0 |
160 |
1 |
kV |
|
|
0 |
10 |
0.1 |
mA |